Set test mode, Set upper limit – LABEC V-18 User Manual
Page 21

16/18 Manual
18
【ENTER】
,
it will return to the kinetics setting interface.
Note:
1. The range of total test time is 1-120,000, other value is invalid.
2.The maximum value of total time is correlative to the interval time,
because system only permits 1000 dots be adopted at most. For
example:
if the interval time is 1s, then the maximum total test time is:
1 X 1000=1000s;
if the interval time is 20s, the maximum total test time is:
20 X 1000=20,000s;
if the interval time is 200s, the maximum total test time is
NOT
:
200 X 1000=200,000s
Because the range of total time is 1-120,000, when the interval time is
200s, the maximum test time is 120,000!
3.
If you don’t want to change the value, press【RETURN】directly.
Set Test Mode
Move the cursor on “
Test Mode”
by pressing
【∧
∧
∧
∧】
and
【∨
∨
∨
∨】
in setting
interface, followed by
【
ENTER
】
.(Fig. 34)
ENTER
RETURN
Fig. 34
Move the cursor on the test mode
by pressing
【∧
∧
∧
∧】
and
【∨
∨
∨
∨】
, then press
【
ENTER
】
to select. Press
【
RETURN
】
to go back to the kinetics setting
interface.
Set Upper Limit
Move the cursor on “
Upper Limit”
by pressing
【∧
∧
∧
∧】
and
【∨
∨
∨
∨】
in setting
interface, followed by
【
ENTER
】
.(Fig. 35)
ENTER
RETURN
Fig. 35
I
nput the new upper value by pressing the numeric keypad followed
by
【ENTER】
,
it will return to the kinetics setting interface.
√ Kinetics
○
○
○
○Time Interval
○
○
○
○Total Time
●
●
●
●Test Mode
√
√
√
√Kinetics
●
●
●
●Abs. √
√
√
√
○
○
○
○T%
○
○
○
○Energy
√ Kinetics
●
●
●
●Upper Limit
○
○
○
○Lower Limit
○
○
○
○Display Data
Kinetics: Abs.
Upper=3.000
Please Input New
Upper=