] filter, ] entries, 4 scale parameters [2--] scale block – Flintec FT-12 Manual User Manual

Page 17: ] set up

background image

FT-12 Technical Manual, Rev. 1.35 November 2010

Page 17 of 44

[105 ] Filling Type (only for filling application)

0 : Gross filling (factory default setting)

1 : Net filling

6.3 Configuration

Parameters

[1--] Configuration Block

[11-] Start Up

[111 X] Auto Power Off

0 : Auto power off disabled (default)

1 : Power off after 3 minutes of stability or no key function

2 : Power off after 10 minutes of stability or no key function

[112 X] Store Tare at Power Off

0 : Disabled (factory default)

1 : Enabled

[113 X] Auto Clear Tare

0 : Disabled (factory default)

1 : The scale returns to gross mode after unloading

[114 X] Buzzer

0 : Disabled (factory default)

1 : The Buzzer will be actived if any key is pressed OR
any error message occurred OR any alarm is active.

[12-] Filter

In this block the proper filter values according to the operating conditions can be entered. One of the most
important features of the indicator is viewing the filter characteristics on the display and with the help of this
option, you can select the most suitable filter without leaving the setup mode.

[120 X ] Filter

The filter value can be set from 0 to 9 (The minimum value of the filter is at least 7 in normal weighing applications).
When you enter this parameter and press the key while [120 X] is shown on the display, the weight
variation can be seen on the display. The value of the filter can be changed by using the <

> and <> keys and the

weight variation for each value is shown on the display. After finalising the filter selection you can go to the next step
by pressing the key.

[14-] Entries

In this block you can enter the initial CN (consecutive number).

[140 ] Date

[DD.MM.YY]

First press the key for the date entry, then enter the new date via numerical keys and
save by pressing .

[141 ] Time

[ HH.MM]

First press the key for the time entry, then enter the new date via numerical keys and
save by pressing .

[142 ] Label No. (CN)

[ XXXXXX ]

The desired value is entered via and keys and saved by pressing . If the
number exceeds 65 535 it will automatically restart at 1.

6.4 Scale

Parameters

[2--] Scale Block

The parameters related to the weighing operation are located here.

[20-] Set Up

[200 X] Approved

0 = No

1 = Yes

[201 X] Increased Indication (10x)

0 = By pressing key

1 = Always increased indication