Flintec FT-11D User Manual
Page 41
FT-11(D) Technical Manual, Rev. 1.35 November 2010
Page 41 of 52
Examples:
The commands will be used for an indicator which instrument address is set to 1.
Reading weight data:
@010300000002FA
Reading status data:
@010300020001F9
Reading Tare data:
@010300030002F7
Taring:
@011000080001020002E2
Loading 1500 value to Setpoint 1:
1.step:
@0110000B000204000005DCFD
2.step:
@011000080001020006DE
Reading Setpoint 1 (SP1=1500):
1. step:
@011000080001020007DD
2. step:
@0103000B0002EF
Break down an example:
Setpoint 1: Write value 1500 into buffer
@
01
10
000B
0002
04
000005DC
FD
STX Device-
Adresse
Write Add.
Reg.
for
Setpoint 1
No. of
Addr. Reg.
No. of
Data Reg.
Value 1500 (dec.),
shown in hex
Checksum EOP
NOTE:
The points you should paid attention to while loading and reading setpoints are:
To write any value to FT-1x the value should be loaded into the buffer (address 000B). Afterwards send
command “6” to address 0008 to make FT-1x to use the value at address 000B.
To read a value from FT-1x you should send command “7” to address 0008 to make FT-1x to load the
value into the buffer. Afterwards you can read the value at address 000B.
The commands “6” and “7” which are sent to address 0008 let you write or read all setpoints