[204 X] Auto Zero Tracking
AZT automatically re-adjusts the scale to zero for compensating defined small deviations around the
centre of zero.
0 = disable
1 = ± 0,5e
2 = ± 1e
3 = ± 3e
[205 X] Tare
0 = Disabled
1 = Multi tare via key
2 = Tare via key in gross mode
[206 X] Motion Detector
This parameter defines the sensitivity level which will determine what is considered as stable.
0 = ± 0.3e
1= ± 0.5e
2= ± 1e
3= ± 2e
4 = No motion detector
[207 X] Tilt Switch (for FT-11 only)
The tilt switch is being used to prevent wrong weighing results in mobile scales due to the slope of the
floor. If this parameter is enabled the digital input 1 must be connected to the tilt switch.
0 = Disabled
1 = Normally Open contact
2 = Normally Closed contact
[208 X] Stability Time
This parameter defines the time period (up to 9.9 s) which will determine what is considered as stable.
[21-] Scale Build
The capacity and the resolution of the scale will be defined here.
[210 X] Scale Type
The capacity and resolution parameters will vary according to the selection here.
0 = Single Range
1 = Dual Range
You will reach the next parameter by pressing
vary according to this selection.
If single range has been selected:
[212 ] Capacity
Press the
[CAP ]
[ XXXXXX ]
The capacity of the scale will be entered here by
The value will be confirmed by pressing
[d ]
[ XXXXXX ]
Display resolution of the scale will be selected by
The selection will be confirmed by pressing
If dual range has been selected:
[211 ] 1. Capacity
Press the
[CAP1 ]
[ XXXXXX ]
The capacity of the first weighing range (Select from Appendix 6) will be entered
here by
[d1 ]
[ XXXXXX ]
Display resolution of the first weighing range will be selected by
confirmed with
[212 ] 2. Capacity
Press the
[CAP2 ]
[ XXXXXX ]
The capacity of the second weighing range (Select from Appendix 7) will be entered
here by
[d2 ]
[ XXXXXX ]
Display resolution of the second weighing range will be selected by
confirmed with
Attention:
After the scale parameters have been changed the scale has to be calibrated.