Continuous output mode, Host mode (for ft-11 only), Print mode – Flintec FT-11D User Manual
Page 23
FT-11(D) Technical Manual, Rev. 1.35 November 2010
Page 23 of 52
The pin configuration of the optional D-Sub 25-pin female connector is given below.
Parameter [012]
Pin no. for stainless steel housing
is set to
“0” or “1”
is set to “2”
Pin no.
for desk and panel types
(D-Sub, 25-pin, female)
J10 connector
(Option 1)
JR1 / JR2 connector
(Option 2 or 3)
TxD1; Par. [01-]
2
8
15
RxD1; Par. [01-]
3
7
14
TxD2; Par. [02-]
RTS1; Par. [01-]
4
3
10
RxD2; Par. [02-]
CTS1; Par. [01-]
5
2
9
GND (RS 232C)
7
6-9
13
R-; Par. [02-]
8
Not used
R+; Par. [02-]
9
Not used
T+; Par. [02-]
10
Not used
+V; Par. [02-]
11
Not used
Termination (RS 485)
19
Not used
Termination (RS 485)
20
Not used
A; Par. [02-]
21
5
12
B; Par. [02-]
22
4
11
T-; Par. [02-]
24
Not used
+24V; 20 mA CL
25
Not used
Shield D25
body
1
1
The RS485 interface termination can be done by short circuiting pin no. 19 and 20 at the desk and panel type
enclosures or by jumper JP4 on the interface board within the stainless steel housing . These short circuits
terminate the line with an internal termination resistor of 100 Ω.
20 mA CL ASCII interface connection:
The 20mA CL ASCII output is disabled by
default. For enabling this output the
connections should be made as given in figure
shown on the left.
(Enabled output is not available for 12 V DC
instruments and stainless steel housings).
8.2.3. Continuous
Output
Mode
The continuous output mode allows for a fast and continuous data export to connected peripheral devices. The
output data structure is described in Appendix 2.
8.2.4.
Host Mode (for FT-11 only)
The indicator can communicate with a PC in host mode. You can upload or download data to the indicator by
adjusting the related interface parameters within the FLINTEC IndFace software (for FT-11 only) or using the
command sets described in Appendix 3. This mode is convenient for process control applications which need
not only the weight data, but also require some uploads / downloads and status control.
8.2.5. Print
Mode
In the print mode you can select the print formats (parameter group [04-], see chapter 8.2.6). This format –
preferably the single line – is convenient to use as weight data interface in legal for trade applications which
require recording to the Alibi memory. The print mode is not available for more than one interface (see also
overview in chapter 8.1).
1. Single Line
You can send the data in single lines as shown below by pressing the
Example from FT-11(D):
CN: 21 G: 3.000kg T: 1.000kg N: 2.000kg
CN
GROSS
TARE
NET
MSD LSD SP MSD LSD SP MSD LSD SP MSD LSD LF CR
9 3 13 3 13 3 13 1
1
25
11
10
24
9
8
FT-13