beautypg.com

EVCO PR100AX9S001 User Manual

Page 3

background image

Every Contr

ol S.r

.l. • PR 100A • Sheet 2/3

6

ALARMS

6.1 Alarms

CODE

AN1...6

Err

probe

1...6

alarm

AN1...6

AH

upper

alarm

probe

1...6

REASONS

• the kind of probe

1...6 you have con-

nected is not right

• the probe 1...6 plays

up

• the connection in-

strument-probe 1...6

is wrong

• the value the probe

1...6 is reading is

outside the limits al-

lowed by the work-

ing range of the in-

strument

• the value the probe

1...6 is reading is

unstable (the value

has changed more

than 1 degree/point

per 2 s for 8 times in

succession)

the value the probe

1...6 is reading is out-

side the limit you have

set with the parameter

AN1...6 Max Alarm

REMEDIES

• look at the param-

eter AN1...6 Type

• test the integrity of

the probe

• test the instrument-

probe connection

• test the value close

to the probe (it has

to be between the

limits allowed by the

working range)

test the value close to

the probe (look at the

parameters AN1...6

Alarm Hyst

and

A N 1 . . . 6 M a x

Alarm

)

EFFECTS

if the On Line Report

printing mode is run-

ning, the instrument

will print and store the

event; if the Daily Re-

port and/or Historical

Report printing modes

are running, the in-

strument will store the

event

if the On Line Report

printing mode is run-

ning, the instrument

will print and store the

event; if the Daily Re-

port and/or Historical

Report printing modes

are running, the inst.

will store the event

7

TECHNICAL DATA

7.1 Technical data

Box: self-extinguishing grey.

Size: 96 x 96 x 90 mm (3.77 x 3.77 x 3.54 in).

Installation: panel mounting, panel cut out 92 x 92 mm (3.62 x 3.62 in), with screw

brackets (they are supplied by the builder).

Frontal protection: IP 30.

Connections: extractable terminal blocks with pitch 5 mm (0.19 in) for cables up to

2.5 mm² (0.38 sq in, inputs and expansion) and with pitch 7.5 mm (0.29 in) for cables

up to 2.5 mm² (0.38 sq in, power supply).

Ambient temperature: from 0 to 55 °C (32 to 131 °F, 10 ... 90% of relative humidity

without condensate).

Power supply: 110-240 Vac, 50/60 Hz (standard) or 12-36 Vac/dc, 50/60 Hz

(by request).

Clock data maintenance without power supply: typically more than 3 years.

Capacity of memory: 2,000 printing lines, independently on the number of measure

inputs (for example, if the acquisition time is 15 min, the capacity of memory will be 500

h that will be about 21 days).

Measure inputs: 2 (one can expand them up to 6 by using the expansion DR 100A)

for PTC or NTC probes, “J” or “K” thermocouples, 2 wires Pt 100 probes, 4-20 mA current

transducers; the probes can be up to 15 m long (49.21 ft).

At terminal 4 there are 12 Vdc you can use in order to supply the transducers.

AN1...6

AL

lower

alarm

probe

1...6

Memory

Full

memory

run out

alarm

the value the probe

1...6 is reading is out-

side the limit you have

set with the parameter

AN1...6 Min Alarm

the memory has run

out

test the value close to

the probe (look at the

parameters AN1...6

Alarm Hyst

and

A N 1 . . . 6 M i n

Alarm

)

erase the data the in-

strument has stored

(look at the parameters

Delete Memory?

and Memory Type)

if the On Line Report

printing mode is run-

ning, the instrument

will print and store the

event;

if the Daily Report

and/or Historical Re-

port printing modes

are running, the in-

strument will store the

event

the instrument will

not store any data

INDICATION

Printing...

Recording...

Memory 90% ... 99%

MEANING

the On Line Report printing mode will be running

the Daily Report and/or Historical Report printing modes

will be running

the memory will be running out

9

ELECTRICAL CONNECTION

9.1 Electrical connection

(19) you have to connect the probes in succession (for example, if you are using three channels, you have to connect probe 1 and probe 2 with the data recorder PR 100A and

probe 3 with the expansion DR 100A); connect PR 100A with DR 100A by using a twisted pair

(20) provide the probe with a protection able to protect it against contacts with metal parts or use insulated probes.