A&D Weighing AD-4407 User Manual
Page 49
AD-4407 Weighing Indicator
Page
47
13.2.
13.2.
13.2.
13.2. Data
Data
Data
Data Format
Format
Format
Format
There are two types of data format set at F-Function f34 and f35.
The initial data format of f34 is shown below.
Header 1 f34, 35 : 9 (Status)
ST Stable weight data.
US Unstable weight data.
OL Overload (Out of range).
HD Holded weight data (The response of the “RW and “RW,n” command).
Header 2 f34, 35 : e (weight type)
GS
or
G
Gross data.
NT
or
N
Net data
TR
or
T
Tare data.
Data
f34, 35 : a, b, C, d (weight data)
The first of the data bits is the polarity, "+" or "-".
When the data is zero, the polarity is "+".
8 digits including polarity and decimal point (dot or comma).
In case of "Out of range", the data are replaced by spaces except the decimal point.
Unit
f34, 35 : F (unit)
In case of Cf07 0, the unit length is 2 digits.
In case of Cf07 1, the unit length is 3 digits. Depending on circumstances, an A&D
printer may not work correctly.
13.3.
13.3.
13.3.
13.3. Command Format
Command Format
Command Format
Command Format
Explanation of
Explanation of
Explanation of
Explanation of Command
Command
Command
Command
When performing a command, the received command or replay data is sent back.
When the received command can not be performed such as the "busy" state, the
code "
I " is sent back. In this case, insert a delay time. Electrical noise may be
the cause of this error.
When receiving an undefined command (unknown command), a "?" is sent back.
13.2. Data Format
Cf06=0,1 Cf06=2
Cf07=0 Cf07=1
2 figures 3 figures
Header 1 Header 2
Unit
8 digit data
Terminators
T
S , S
G
+
,
0
0
0
0 0
0
.
g
k
T
S
S
U
L
O
D
H
G
N
T
T
P
S
G
T
N
R
T
T
P
g
C
P
t
g
k
g
t
C
P
g
k
l
t
b
l
l
t
b
l
z
o
y
c
z
o
y
c