Cf-functions – A&D Weighing AD-4407 User Manual
Page 32
![background image](/manuals/659075/32/background.png)
Page
30
AD-4407 Weighing Indicator
6.3. CF-Functions
Item
Parameter
Description
* 0
No limitation
Cf00
Zero track width, motion
detection condition
1
Use limitation at f01, f02, f03, f27 and f28.
* 0
2% of CAP, Tare limit is 100 % CAP
1
10% of CAP, Tare limit is 100 % CAP
2
3% of CAP, Tare limit is 50 % CAP
Cf01
Zero range
Turning display on, the
range to zero display.
3
4% of CAP, Tare limit is 50 % CAP
0
Not to zero when turning the display on.
* 1
10% of CAP
2
3% of CAP
Cf02
Power on zero range
Turning display on, the
range to zero display.
3
4% of CAP
0
Gross when displaying gross.
1
Gross
Cf03
Zero tracking
* 2
Gross or Net when displaying net.
TARE, ZERO in motion / TARE at negative gross
* 0
Not accepted
/
Not accepted
1
Execute /
Not
accepted
2
Not accepted
/
Execute
Cf04
TARE, ZERO in motion /
TARE at negative gross
value
3
Execute /
Execute
* 0
Not to output data at unstable value or over load.
Effective in key mode.
Cf05
Output on over load and
unstable state.
1
To output data always.
GROSS / NET / TARE / Preset TARE
* 0
GR / NT / TR /
TR
1
GR / NT / TR /
PT
Cf06
Header 2
2
G_
/ N_ / T_ /
PT (_:Space 20h)
* 0
Two digits
Cf07
Figure number of unit
1
Three digits
Serial output data
* 0
Not used (Ineffective)
Cf08
Accumulation function
1
Used (Effective)
* 0
Used (Effective)
Cf09
Digital Tare(PT)
1
Not used (Ineffective)
CAP : maximum capacity
*: Initial settings
Power on ZERO (CF02)
Cf02 0
After power on, weight display starts immediately.
Cf02 1,2,3
When the zero range is exceeded at power on, ------ is displayed.
Press
the
ESC
key to start weight display.
6.3. CF-Functions