AD-4407 Weighing Indicator
Note that the
ON/OFF
key does not function alone. Press the
ESC
key while
holding the
ON/OFF
key to end the calibration mode, if mis-operation. After
displaying Can5el, press the
ON/OFF
key to stop calibration mode and to
turn the indicator off.
NOTE: When displaying Caloff , press the +/- key while pressing the
ON/OFF
key,
instead of
CAL
key.
Caution
The maximum display is less than or equal to 40000 divisions. This
number is calculated from the maximum capacity divided by the
minimum division.
Check the accuracy of weighing instrument periodically.
Recommended mass, use a mass heavier than 2/3 maximum capacity.
Calibrate the scale, if it is moved to other location or the environment
has changed.
It is not necessary to input the gravity acceleration correction, when
calibrating the scale with a calibration mass at the place where the
scale is used.
Enter the stable weighing data while the STABLE mark is displayed. If
unstable data is used, it may cause a weighing error. Arrange the
condition using the F00 filter function.
The span calibration needs the zero calibration data. We recommend
that you perform the span calibration immediately after the zero
calibration.
If you use the dual range function of the multi-interval scale, perform
the "Range Function", "Zero Calibration" and "Span Calibration".
5.2.
5.2.
5.2.
5.2. Calibration
Procedure
Calibration Procedure
Calibration Procedure
Calibration Procedure
5.2.1.
Configuring a Weighing Instrument
This section explains how to specify capacity, resolution, decimal point position and
display format, weighing range and unit. Perform this procedure when installing the
indicator.
When Cal5et is displayed, enter this mode by pressing of the ENTER key.
Specify the range and unit.