AD-4407 Weighing Indicator
Page
13
Single Range
Select resolution and decimal point position and format.
Specify the weighing capacity.
Dual Range
For the range function, refer to “5.3. Weighing Range Function”.
Specifying the Range and Unit
Specifying the Range and Unit
Specifying the Range and Unit
Specifying the Range and Unit
Step 1
The range and unit of measure are displayed.
Range display : 5ingl : single range
dUal : dual range
To change the range function, use the
CLEAR
key
Unit display:
The active unit is displayed. The unit can be changed such as
kg or lb. The calibration is performed with displayed unit.
Press
the
MODE
key to select a unit and to select a unit for calibration,
use the +/- key. The unit for calibration (first unit) is displayed and the
alternate unit (second unit) is blinking.
ENTER
The key to store the data displayed and proceed to the next step.
ESC
The key to proceed to the next step without changing them.
Specifying the Resolution, Decimal Point Position an
Specifying the Resolution, Decimal Point Position an
Specifying the Resolution, Decimal Point Position an
Specifying the Resolution, Decimal Point Position and Format
d Format
d Format
d Format
Step 2 The resolution will be displayed as d 0.1 with decimal point. The indicator displays
triangle
1 and the first unit selected at the previous step. Specify the position of
the decimal point with the
MODE
key. Specify the display format (point or comma)
with the +/- key. Specify the resolution with
CLEAR
key. The decimal point format
is specified in step. The decimal point format for serial data output is selected
using the F-function settings. Press the
ENTER
key to store them and proceed to
the next step. When pressing the
ESC
key, regardless of what is displayed, the
indicator will proceed to the next step without changing them.
5.2. Calibration Procedure