A&D Weighing AD-4407 User Manual
Page 34
Page
32
AD-4407 Weighing Indicator
8.
8.
8.
8. Accumulation
Accumulation
Accumulation
Accumulation
The function accumulates weighing data and stores the total data and the
accumulation count. Data is stored in non-volatile memory, and is not lost even if
the power is turned off.
8.1.1. Preparation
and
Specification
Set the following parameters to use the accumulation function.
Select
Cf08 1 for the CF-function so that the accumulation function becomes effective.
Specify the method of accumulation and data at f20 of the F-functions.
Specify the inhibit region for accumulation at f21 of the F-functions.
Selection of Accumulation Mode, F20 of the F-functions
There are two methods of accumulation; manual accumulation using the
M+
key
and automatic accumulation.
The accumulated data can select "positive data only" or "both polarity data".
The accumulated data can select "result of comparison is OK only" or "all result".
Accumulation Condition, F21 of the F-function
In the case of manual accumulation mode, press the
M+
key to accumulate
weighing data when the STABLE mark is turned on.
Data can be accumulated after the weight data enters the "inhibit region for output".
When connecting the power cord and turning the display on, the accumulation mode
takes the same action.
Inhibit region for accumulation
F21
Description
Add data anytime
f21 0
Stable data can be used anytime
Above 5 d
f21 1
Factory setting
Above 10 d
f21 2
Above 20 d
f21 3
Above 50 d
f21 4
Caution
Do not set to
f21 0
f21 0
f21 0
f21 0 for the automatic accumulation mode.
If
f21 0
f21 0
f21 0
f21 0, it may add the same data two times or more.
Limitation of Accumulation Count and Total
The limitation of accumulation count is 999999.
The limitation of total is 999999, ignoring the decimal point.
If exceeding these limitations, the data is not accumulated.
Example: The decimal point to "0.0", the limitation is "99999.9".
8. Accumulation