5. calibration calibration calibration calibration – A&D Weighing AD-4407 User Manual
Page 13
AD-4407 Weighing Indicator
Page
11
5.
5.
5.
5. Calibration
Calibration
Calibration
Calibration
This indicator, converts an input voltage from a load cell to the "mass" value, and
displays it. Calibration is the adjustment function so that the scale (indicator) can
display the weight correctly.
5.1.1.
Items of Calibration Mode
There are four items in the calibration function in the procedure.
How to calibrate: In weighing mode, press the CAL key. After Cal in is displayed
for 2 seconds Cal 0 will appear. Then the required items should be selected and
displayed with the
MODE
key, then executed by pressing
ENTER
key.
NOTE: Calibration could be started by simultaneously pressing the
ZERO
and
TARE
key,
instead of CAL key. However, in some cases, the procedures are altered to nullify this
option. Furthermore, this option is not available with Stamping type version setting.
Required Items
Cal 5et
Store capacity, resolution, alignment of decimal point position and
display format, weighing range and unit. These items should be input
first in order for the indicator to function as a weighing instrument.
These parameters do not need to be changed again unless the
indicator itself is replaced. For details, refer to “5.2.1.Configuring a
Weighing Instrument”.
Cal 0
Calibrates zero and span. This is required after installation, to get
accurate data. For details, refer to “5.2.3. Zero Calibration” and “5.2.4.
Span Calibration”.
Optional Items (Sub-functions)
lnr 0
Performs digital linearization. Refer to “5.4.Digital Linearization
Function”.
g 5et
Compensates for acceleration of gravity.
Refer to “5.5. Gravity
Compensation Function”.
Gravity compensation function: Compensates for weighing error
between the calibration location and other weighing location using
gravity acceleration.
In the calibration mode the keys have functions as follows:
0
-
9
Numerical
keys.
MODE
The key to display other items.
CLEAR
The cancel key at inputting data, initial data and changing mode.
+/-
The key to display other parameters.
ESC
The key to proceed to the next step without changing them.
ENTER
The key to store new calibration data and proceeds to the next step.
CAL
The key to store all data into memory and display Caloff after the
calibration mode. And press the
ON/OFF
key to turn off the display.
5.
Calibration