beautypg.com

KEYENCE N-R2 User Manual

N-r2, Instruction manual, Dedicated communication unit rs-232c output type

background image

1

E N-R2-IM

Dedicated Communication Unit
RS-232C Output Type

N-R2

Instruction Manual

General precautions

At startup and during operation, be sure to check that the
functions and performance of the N-R2 is operating properly.

We recommend that you take substantial safety measures to
avoid any damage in the event a problem occurs.

If the N-R2 is modified or used in any way other than described in
the specifications, its function and performance cannot be
guaranteed.

When the N-R2 is used in combination with other instruments, the
function and performance may be effected by operating conditions
and the surrounding environment.

Do not use the N-R2 for the purpose of protecting the human
body.

The information contained in this manual is subject to change
without notice.

Notice

When this product is used under the circumstances and operating
environments described below, adhere to the limitations of the ratings,
take adequate measures to ensure safety such as fail-safe installations
and consult a KEYENCE sales representative.

For use under circumstances or environments which are not
described in the manual

For use with nuclear power control, railway, aircraft, vehicles,
incinerators, medical equipment, entertainment equipment, safety
devices etc.

For use in applications where death or serious property damage is
possible and extensive safety precautions are required.

For properly using the product

Do not install the N-R2 in the following locations:

Locations subject to direct sunlight

Locations where the surrounding air temperature is outside the
range of 0

to +50

°C or locations subject to condensation as the

result of severe changes in temperature

Locations where the ambient humidity is outside the range of 35
to 85%RH.

Locations subject to corrosive or flammable gases

Locations subject to dust, salts, iron dust, or oil smoke

Locations subject to shock or vibration

Locations subject to exposure of water, oil, or chemicals

Locations where strong magnetic or electric fields are generated

UL certification

The N-R2 complies with the following UL/CSA standards and has
obtained the UL/C-UL certification.

Applicable standards

UL508, UL60950-1

UL File No.

E207185, E167973

UL category

NRAQ/NRAQ7, NWGQ2/NWGQ8

Be sure to observe the following installation and environment conditions.

Pollution degree 2

Overvoltage category I

Caution - Do not connect directly to the branch circuit. This product must
be supplied power by a suitable, approved isolated transformer or power
supply not exceeding 200 VA max.

Attention - Ne raccordez pas l'unité directement au circuit de dérivation.
Ce produit doit être alimenté par un transformateur adapté, isolé et
approuvé ou une alimentation ne dépassant pas 200 VA.

FCC regulations

The N-R2 complies with the following FCC EMI regulations.

FCC Part 15 Subpart B, Class B digital device

Canada IC (Industry Canada) regulations

The N-R2 complies with the following IC EMI regulations.

ICES-003, Class B digital apparatus

CE marking

The N-R2 complies with the following essential requirements of the EU
Directive.

EMC Directive (2004/108/EC)

The following harmonized standards are applied to this product in order
to confirm the compliance.

Applicable standards (EMI)

EN55011, Class A
EN55022, Class B

Applicable standards (EMS)

EN61000-6-2
EN61000-6-1

* The length of the communication and power cables should be less than 30 m.

DANGER

Failure to follow these instructions may lead to death or
serious injury.

WARNING

Failure to follow these instructions may lead to injury.

CAUTION

Failure to follow these instructions may lead to product
damage (product malfunction, etc.).

Note

Provides additional information on proper operation.

TIPS

Indicates useful information or information that aids
understanding of text descriptions.

For Safely Using the Product

CAUTION

Do not use the N-R2 in an environment subject to

flammable, explosive, or corrosive gases.

Be sure that all screws are tightened securely.
Use crimp contacts of the specified size for wiring.
The N-R2 uses a 24 VDC power supply. Using a power

supply outside this range or an AC power supply may
cause product failure.

Do not disassemble or modify the N-R2. This may cause

product failure.

Locate the cables as far as possible from high-voltage

lines and power lines. Otherwise, generated noise may
cause product failure or malfunctions.

The N-R2 is a precision instrument. If the unit is

dropped or shock is applied, it may be damaged. Use
caution when unpacking, carrying, and mounting.

Be sure to observe all warnings, cautions, and

precautions specified in the manual.

96M13455

Foreign Regulations and Standards

Package Contents List

Main unit

Instruction manual

1

E N-R2-IM

Dedicated Communication Unit
RS-232C Output Type

N-R2

Instruction Manual

General precautions

At startup and during operation, be sure to check that the
functions and performance of the N-R2 is operating properly.

We recommend that you take substantial safety measures to
avoid any damage in the event a problem occurs.

If the N-R2 is modified or used in any way other than described in
the specifications, its function and performance cannot be
guaranteed.

When the N-R2 is used in combination with other instruments, the
function and performance may be effected by operating conditions
and the surrounding environment.

Do not use the N-R2 for the purpose of protecting the human
body.

The information contained in this manual is subject to change
without notice.

Notice

When this product is used under the circumstances and operating
environments described below, adhere to the limitations of the ratings,
take adequate measures to ensure safety such as fail-safe installations
and consult a KEYENCE sales representative.

For use under circumstances or environments which are not
described in the manual

For use with nuclear power control, railway, aircraft, vehicles,
incinerators, medical equipment, entertainment equipment, safety
devices etc.

For use in applications where death or serious property damage is
possible and extensive safety precautions are required.

For properly using the product

Do not install the N-R2 in the following locations:

Locations subject to direct sunlight

Locations where the surrounding air temperature is outside the
range of 0

to +50qC or locations subject to condensation as the

result of severe changes in temperature

Locations where the ambient humidity is outside the range of 35
to 85%RH.

Locations subject to corrosive or flammable gases

Locations subject to dust, salts, iron dust, or oil smoke

Locations subject to shock or vibration

Locations subject to exposure of water, oil, or chemicals

Locations where strong magnetic or electric fields are generated

UL certification

The N-R2 complies with the following UL/CSA standards and has
obtained the UL/C-UL certification.

Applicable standards

UL508, UL60950-1

UL File No.

E207185, E167973

UL category

NRAQ/NRAQ7, NWGQ2/NWGQ8

Be sure to observe the following installation and environment conditions.

Pollution degree 2

Overvoltage category I

Caution - Do not connect directly to the branch circuit. This product must
be supplied power by a suitable, approved isolated transformer or power
supply not exceeding 200 VA max.

FCC regulations

The N-R2 complies with the following FCC EMI regulations.

FCC Part 15 Subpart B, Class B digital device

Canada IC (Industry Canada) regulations

The N-R2 complies with the following IC EMI regulations.

ICES-003, Class B digital apparatus

CE marking

The N-R2 complies with the following essential requirements of the EU
Directive.

EMC Directive (2004/108/EC)

The following harmonized standards are applied to this product in order
to confirm the compliance.

Applicable standards (EMI)

EN55011, Class A
EN55022, Class B

Applicable standards (EMS)

EN61000-6-2
EN61000-6-1

* The length of the communication and power cables should be less than 30 m.

DANGER

Failure to follow these instructions may lead to death or
serious injury.

WARNING

Failure to follow these instructions may lead to injury.

CAUTION

Failure to follow these instructions may lead to product
damage (product malfunction, etc.).

Note

Provides additional information on proper operation.

TIPS

Indicates useful information or information that aids
understanding of text descriptions.

For Safely Using the Product

CAUTION

Do not use the N-R2 in an environment subject to

flammable, explosive, or corrosive gases.

Be sure that all screws are tightened securely.
Use crimp contacts of the specified size for wiring.
The N-R2 uses a 24 VDC power supply. Using a power

supply outside this range or an AC power supply may
cause product failure.

Do not disassemble or modify the N-R2. This may cause

product failure.

96M00266

CAUTION

Locate the cables as far as possible from high-voltage

lines and power lines. Otherwise, generated noise may
cause product failure or malfunctions.

The N-R2 is a precision instrument. If the unit is

dropped or shock is applied, it may be damaged. Use
caution when unpacking, carrying, and mounting.

Be sure to observe all warnings, cautions, and

precautions specified in the manual.

Foreign Regulations and Standards

Package Contents List

Main unit

Instruction manual

1

E N-R2-IM

Dedicated Communication Unit
RS-232C Output Type

N-R2
Instruction Manual

General precautions

At startup and during operation, be sure to check that the
functions and performance of the N-R2 is operating properly.

We recommend that you take substantial safety measures to
avoid any damage in the event a problem occurs.

If the N-R2 is modified or used in any way other than described in
the specifications, its function and performance cannot be
guaranteed.

When the N-R2 is used in combination with other instruments, the
function and performance may be effected by operating conditions
and the surrounding environment.

Do not use the N-R2 for the purpose of protecting the human
body.

The information contained in this manual is subject to change
without notice.

Notice
When this product is used under the circumstances and operating
environments described below, adhere to the limitations of the ratings,
take adequate measures to ensure safety such as fail-safe installations
and consult a KEYENCE sales representative.

For use under circumstances or environments which are not
described in the manual

For use with nuclear power control, railway, aircraft, vehicles,
incinerators, medical equipment, entertainment equipment, safety
devices etc.

For use in applications where death or serious property damage is
possible and extensive safety precautions are required.

For properly using the product

Do not install the N-R2 in the following locations:

Locations subject to direct sunlight

Locations where the surrounding air temperature is outside the
range of 0

to +50qC or locations subject to condensation as the

result of severe changes in temperature

Locations where the ambient humidity is outside the range of 35
to 85%RH.

Locations subject to corrosive or flammable gases

Locations subject to dust, salts, iron dust, or oil smoke

Locations subject to shock or vibration

Locations subject to exposure of water, oil, or chemicals

Locations where strong magnetic or electric fields are generated

UL certification
The N-R2 complies with the following UL/CSA standards and has
obtained the UL/C-UL certification.

Applicable standards

UL508, UL60950-1

UL File No.

E207185, E167973

UL category

NRAQ/NRAQ7, NWGQ2/NWGQ8

Be sure to observe the following installation and environment conditions.

Pollution degree 2

Overvoltage category I

Caution - Do not connect directly to the branch circuit. This product must
be supplied power by a suitable, approved isolated transformer or power
supply not exceeding 200 VA max.

FCC regulations
The N-R2 complies with the following FCC EMI regulations.

FCC Part 15 Subpart B, Class B digital device

Canada IC (Industry Canada) regulations
The N-R2 complies with the following IC EMI regulations.

ICES-003, Class B digital apparatus

CE marking
The N-R2 complies with the following essential requirements of the EU
Directive.

EMC Directive (2004/108/EC)

The following harmonized standards are applied to this product in order
to confirm the compliance.

Applicable standards (EMI)

EN55011, Class A
EN55022, Class B

Applicable standards (EMS)

EN61000-6-2
EN61000-6-1

* The length of the communication and power cables should be less than 30 m.

DANGER

Failure to follow these instructions may lead to death or
serious injury.

WARNING

Failure to follow these instructions may lead to injury.

CAUTION

Failure to follow these instructions may lead to product
damage (product malfunction, etc.).

Note

Provides additional information on proper operation.

TIPS

Indicates useful information or information that aids
understanding of text descriptions.

For Safely Using the Product

CAUTION

Do not use the N-R2 in an environment subject to
flammable, explosive, or corrosive gases.
Be sure that all screws are tightened securely.
Use crimp contacts of the specified size for wiring.
The N-R2 uses a 24 VDC power supply. Using a power
supply outside this range or an AC power supply may
cause product failure.
Do not disassemble or modify the N-R2. This may cause
product failure.

96M00266

CAUTION

Locate the cables as far as possible from high-voltage
lines and power lines. Otherwise, generated noise may
cause product failure or malfunctions.

The N-R2 is a precision instrument. If the unit is
dropped or shock is applied, it may be damaged. Use
caution when unpacking, carrying, and mounting.

Be sure to observe all warnings, cautions, and
precautions specified in the manual.

Foreign Regulations and Standards

Package Contents List

Main unit

Instruction manual