Compex Energy is designed for use in the electromagnetic environment stipulated
below. The buyer or user of the Compex Energy must ensure it is used in this
recommended environment.
RECOMMENDATIONS AND DECLARATION BY THE MANUFACTURER -
NOTE: VT is the AC supply voltage before application of the test
level.
Immunity test
Test level IEC 60601
RF immunity
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
10 V/m
26 MHz to 1 GHz
Observance level
Signal lines
Not applicable for power supplies
of battery-powered systems
3 A/m
10 V/m
NOTE 1: From 80 MHz to 800 MHz, the spacing for high frequency amplitude is applied.
NOTE 2: These guidelines may not be appropriate for some situations.
a
The field intensity from fixed transmitters, such as radio telephone base stations (cellular/wireless) and a
mobile radio, amateur radios, AM and FM radio emissions and TV emissions cannot be predicted with any
accuracy. It may therefore be necessary to consider an analysis of the electromagnetic environment of the
site to calculate the electromagnetic environment coming from RF fixed transmitters. If the field intensity
measured in the environment where the Compex Energy is located exceeds the appropriate RF observance
level above, the Compex Energy should be monitored to ensure it is operating properly.