English – Compex Energy mi-Ready User Manual
Page 179
![background image](/manuals/656425/179/background.png)
177
English
0,01
0,12
0,1
0,38
1
1,2
In the case of transmitters whose maximum output power is not shown in the table above, the recommended
spacing of d metres (m) can be calculated using the appropriate equation for the transmitter frequency, where
P is the maximum output power of the transmitter in watts (W) as set by the transmitter manufacturer.
The Compex Energy is designed for use in an electromagnetic environment in which
radiated RF waves are controlled. The buyer or user of the Compex Energy can
contribute to preventing electromagnetic interference by maintaining a minimum
distance between RF portable and mobile communication appliances (transmitters) and
the et Compex Energy according to the table of recommendations below and according
to the maximum output power of the telecommunication appliance.
RECOMMENDED SPACING BETWEEN A PORTABLE AND MOBILE
COMMUNICATION APPLIANCE AND THE COMPEX ENERGY
NOTE 1: At 80 MHz and at 800 MHz, the spacing
for high frequency amplitude is applied.
NOTE 2: These guidelines may not be appropriate for some
situations. Electromagnetic wave propagation is modified by
absorption and reflection due to buildings, objects and persons.
0,12
0,23
0,38
0,73
1,2
2,3
10
3,8
100
12
3,8
7,3
12
23
Maximum transmitter
output power W
From 150 kHz
to 80 MHz
d = 1.2 √P
From 80 MHz
to 800 MHz
d = 1.2 √P
From 800 MHz
to 2.5 GHz
d = 2.3 √P
Spacing according to the frequency of the transmitter mCISPR 11
ENERGY2 FR DE IT ES EN:ENGY Man FR 31/01/08 14:36 Page 177