Compex Energy is designed for use in the electromagnetic environment
stipulated below.
The buyer or user of the Compex Energy must ensure it is used
in this recommended environment.
RECOMMENDATIONS AND DECLARATION BY THE MANUFACTURER -
NOTE: VT is the AC supply voltage before application of the test
level.
Immunity test
Test level IEC 60601
Electrostatic discharge
(DES)
CEI 61000-4-2
± 6 kV at the contact
± 8 kV in air
Rapid electrical transients in
bursts of noise
CEI 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
Surge currents
CEI 61000-4-5
± 1 kV differential mode
± 2 kV common mode
Voltage trough, short power cuts
and voltage variations on power
supply input lines
CEI 61000-4-11
< 5% VT (trough > 95% of UT)
for 0.5 cycle
< 40% VT (trough > 60% of VT)
for 5 cycles
< 70% VT (trough > 30% of VT)
for 25 cycles
< 5% VT (trough > 95% of VT)
for 5 seconds
Observance level
± 6 kV at the contact
± 8 kV in air
Not applicable
System battery-powered
Not applicable
System battery-powered
Not applicable
System battery-powered
Magnetic field
at mains frequency
(50/60 Hz)
CEI 61000-4-8
3 A/m