Emissions test
Compliance
Electromagnetic environment - guide
CISPR 11
RF emissions
Group 1
The Compex Energy uses RF energy only for its
internal operation. Consequently, its RF emissions
are very low and are unlikely to interfere
with any adjacent electrical device.
CISPR 11
RF emissions
Class B
The Compex Energy is suitable for use in any
establishment, including a private dwelling and a place
connected directly to the low voltage mains supply
which powers residential buildings.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage
fluctuations/emission
oscillations
IEC 61000-3-3
Not applicable
The Compex Energy is intended for use in the electromagnetic
environment specified below.
The customer or user of the Compex Energy should ensure
that it is used in this environment.