Model 5i Digital Force / Torque Indicator

User’s Guide


24

8. After displaying “PLEASE WAIT…” the display will appear as follows:

Remove the load applied in Step 8, leave the fixtures in place, then press ZERO.

9. The display will appear as follows:

Use the UP and DOWN keys to adjust the load value as required. The load values default to even
increments, as indicated by the previously entered number of data points (even increments are
recommended for best results). For example, if a 50 lbF capacity sensor is calibrated, and 5 data
points were selected, the load values will default to 10, 20, 30, 40, and 50 lb. Apply the calibration
load. Then press ENTER.

Repeat the above step for the number of data points selected.

10. After all the compression calibration points have been completed, the display will appear as

follows:

Press ENTER.

11. At the completion of the tension calibration, the display will appear as follows:

To save the calibration information, select “SAVE & EXIT”. To exit without saving the data select
“EXIT W/O SAVING”.

CALIBRATION

COMPRESSION

ENSURE NO LOAD

THEN PRESS ZERO

CALIBRATION

COMPLETE


SAVE & EXIT

EXIT W/O SAVING

CALIBRATION

COMPRESSION COMPLETE

REVERSE DIRECTION
FOR TENSION

Attach necessary
weight fixtures.

THEN PRESS ENTER

CALIBRATION

COMPRESSION

APPLY LOAD

1 OF 5

ENTER LOAD:

2.000 LBF

THEN PRESS ENTER