Checkline Series-5I User Manual
Page 20
![background image](/manuals/656229/20/background.png)
Model 5i Digital Force / Torque Indicator
User’s Guide
19
Three options are available:
Selection
Function when pressing DATA
RS232/USB Output
Outputs data via the serial and USB ports
Mitutoyo Output
Outputs data via Mitutoyo (Digimatic) through the serial port
Memory Storage
Stores a reading to memory (refer to the Memory section for details)
Any combination of the above functions may be selected.
9.7 I/O Connector Pin Diagram (female)
Pin No.
Description
Input / Output
1 Signal
Ground
---
2
Tension/Counter-clockwise Overload
Output
3
RS-232 Receive
Input
4 RS-232
Transmit
Output
5 +12V
DC
Output
6 Analog
Output
Output
7 Compression/Clockwise
Overload
Output
8 Mitutoyo
Clock
Output Bit 2
Output
9 Mitutoyo
Data
Output Bit 0
Output
10 Mitutoyo
Request
Input Bit 3
Input
11
“Under” Set Point
Output
12
“Over” Set Point
Output
13
“Within” Set Point
Output
14 External
Trigger
Input
15 Mitutoyo
Ready
Output Bit 1
Output
9.8 Command Set / Gauge Control Language 2 (GCL2)
The 5i may be controlled by an external device through the RS-232 or USB channel. The following is a list
of supported commands and their explanations. All commands must be terminated with a Carriage Return
character or with a Carriage Return/Line Feed combination. The indicator responses are always
terminated with a Carriage Return/Line Feed.
Request
Readings
?
Request the displayed reading (dependant on operating mode)
?C
Request the current (real time) reading
?PT
Request the peak tension reading
?PC
Request the peak compression reading
?CW
Request the peak clockwise reading
?CCW
Request the peak counter-clockwise reading
?ET
Request the reading obtained during the External trigger mode