beautypg.com

Checkline Series-5I User Manual

Page 13

background image

Model 5i Digital Force / Torque Indicator

User’s Guide


12


Two filters are available:

Current Reading – Applies to the peak capture rate of the instrument.

Displayed Reading – Applies to the primary reading on the display.

Available settings: 1,2,4,8,16,32,64,128,256,512,1024. It is recommended to keep the current reading
filter at its lowest value for best performance, and the displayed reading filter at its highest value for best
stability.

7 SET POINTS


7.1 General Information

Set points are useful for tolerance checking (pass/fail),
triggering an external device such as a motorized test
stand, or alarm indication in process control applications.
Two limits, high and low, are specified and stored in the
non-volatile memory of the instrument and the primary
reading is compared to these limits. The results of the
comparisons are indicated through the three outputs
provided on the 15-pin connector, thus providing “under”,
“in range”, and “over” signaling. These outputs can be
connected to indicators, buzzers, or relays as required
for the application.



7.2 Configuration
To configure set points, select Set Points from the menu. The screen will appear as follows:


Either one, two, or none of the set points may be enabled. To toggle between the tension and
compression (or clockwise and counter-clockwise) directions, press the DIRECTION key.

If two set points have been enabled, they are displayed in the upper left corner of the display. If only one
set point has been enabled, the word “OFF” will appear in place of the value. If no set points have been
enabled, the upper left corner of the display will be blank.

DIGITAL FILTERS

(1 = Fastest)

Current Reading

1

Displayed Reading
1

SET POINTS

Upper Disabled
* Upper Enabled

5.000

Lower Disabled
* Lower Enabled

3.500