Checkline Series-5I User Manual
Page 24
Model 5i Digital Force / Torque Indicator
User’s Guide
23
4. Place the sensor horizontally on a level surface free from vibration, then press ZERO. The
indicator will calculate offsets, and the display will appear as follows:
If failed:
5. The following screen appears after the offsets have been calculated:
Attach weight fixtures (brackets, hooks, etc), as required. Do not yet attach any weights or apply
any calibration loads. Then press ENTER.
6. The display will appear as follows:
Optionally exercise the sensor several times (at full scale, if possible), then press ENTER.
7. The display will appear as follows:
Apply a weight equal to the full scale of the instrument, then press ENTER.
CALIBRATION
COMPRESSION
GAIN ADJUST
APPLY FULL SCALE LOAD
10.000 LBF +/-20%
THEN PRESS ENTER
CALIBRATION
COMPRESSION
Attach necessary
weight fixtures.
THEN PRESS ENTER
CALIBRATION
COMPRESSION
Optionally exercise
load cell a few times.
THEN PRESS ENTER
CALIBRATION
OFFSET
Sen.Offset Adj.Failed
Ana.Offset Adj.Failed
CALIBRATION
OFFSET
Sen.Offset Adj.Passed
Ana.Offset Adj.Passed
CALIBRATION
OFFSET
Please wait…