Checkline Series-5I User Manual
Page 21

Model 5i Digital Force / Torque Indicator
User’s Guide
20
?A
Request the average reading obtained during the Average mode
Units (available units depend on the sensor used)
LB
Switch unit to pound-force
OZ
Switch unit to ounce-force
KG
Switch unit to kilogram-force
G
Switch unit to gram-force
N
Switch unit to Newton
MN
Switch unit to milli-Newton
KN
Switch
unit to kilo-Newton
LBFT
Switch unit to pound-foot
LBIN
Switch unit to pound-inch
OZIN
Switch unit to ounce-inch
KGM
Switch unit to kilogram-meter
KGMM
Switch unit to kilogram-millimeter
GCM
Switch
unit
to gram-centimeter
NM
Switch unit to Newton-meter
NCM
Switch unit to Newton-centimeter
NMM
Switch unit to Newton-millimeter
Basic Functions (available measurement directions depend on the sensor used)
CUR
Current mode (real time mode) for primary reading
PT
Peak Tension mode for primary reading
PC
Peak Compression mode for primary reading
PCW
Peak Clockwise mode for primary reading
PCCW
Peak Counter-clockwise mode for primary reading
CLR
Clear peaks
Z
Zero display and perform the CLR function
Filters
FLTCn
Digital filter for displayed readings
FLTPn
Digital filter for current readings
n= 0-10, filter = 2
n
, ex: n=0= no filter, n=10=1024 samples averaged
Memory & Statistics
MEM
Transmit all stored readings
STA
Transmit
statistics
Set
Points
SPHD
Disable high set point
SPLD
Disable low set point
SPHn
High set point. n=value (+ for compression, - for tension)
SPLn
Low set point. n=value (+ for compression, - for tension)
Note: High set point value must be greater than low set point value if both values
are set to the same polarity.
USB/RS-232
Communication
FULL
USB/RS-232 transmission with units
NUM
USB/RS-232 transmission without units (only numeric values)
AOUTn
Auto-transmit n times per second n=1,2,5,10,25,50,125,250. 0=disabled
Note: n = 1 = yields 50 times per second. This is provided for backward
compatibility with the legacy indicator model BGI.
Mitutoyo
Communication
MIT
Enable Mitutoyo output
MITD
Disable Mitutoyo output