Series 5 Digital Force Gauges

User’s Guide


22

Remove the load applied in Step 8, leave the fixtures in place, then press ZERO.

10. The display will appear as follows:

Use the UP and DOWN keys to adjust the load value as required. The load values default to even
increments, as indicated by the previously entered number of data points (even increments are
recommended for best results). For example, if a 50 lbF capacity gauge is calibrated, and 5 data
points were selected, the load values will default to 10, 20, 30, 40, and 50 lb. Apply the calibration
load. Then press ENTER.

Repeat the above step for the number of data points selected.

11. After all the compression calibration points have been completed, the display will appear as

follows:

Press ENTER.

12. The display will appear as follows:

Reverse the orientation of the load cell shaft by rotating the gauge 180 degrees. Press
DIRECTION to invert the display. Then attach weight fixtures. The following screens will step
through the same procedure as with the compression direction. Proceed in the same manner.

CALIBRATION

To invert the
display, press the
DIRECTION button.

THEN PRESS ENTER

CALIBRATION

COMPRESSION

ENSURE NO LOAD

THEN PRESS ZERO

CALIBRATION

COMPRESSION COMPLETE

REVERSE DIRECTION
FOR TENSION

Attach necessary
weight fixtures.

THEN PRESS ENTER

CALIBRATION

COMPRESSION

APPLY LOAD

1 OF 5

ENTER LOAD:

2.000 LBF

THEN PRESS ENTER