beautypg.com

Checkline Series-5 User Manual

Page 20

background image

Series 5 Digital Force Gauges

User’s Guide


19

External Trigger

ETH

Enable high level-triggered External trigger mode

ETL

Enable low level-triggered External trigger mode

ETHL

Enable reading captured on a high to low transition

ETLH

Enable reading captured on a low to high transition

ETD

Disable External trigger mode


Input / Output Bits

Sn

Set output bit (open drain, pull to ground). n=0,1,2

Cn

Clear output bit. n=0,1,2

Rn

Read current status of output bit or level of input pin. n=0,1,2,3

Personality

RN

Read product name

RM

Read

model

number

RV

Read firmware version number

RS

Read serial number

Other

Commands

AOFFn

Auto-shutoff. n=0-30 minutes. 0=auto shutoff disabled

SAVE

Save current settings in nonvolatile memory

LIST

List current settings and status


Following is an example LIST output: V1.00;LBF;CUR;FLTC8;FLTP1;AOUT00;AOFF5;FULL;MIT;POL;B0

All fields are separated by “;”. The first field shows the firmware version, the last field shows the remaining
battery power (B0=full charge, B3=minimum power). All other fields show the status of settings and
features using the same abbreviations as the commands to set them.

Any detected errors are reported back by means of the following error codes:

*10

Illegal

command

*11

Not

applicable

*12

Invalid

specifier

*22 Value too large

10 CALIBRATION

10.1 Initial Physical Setup

The gauge should be mounted vertically to a test stand or fixture rugged enough to withstand a load
equal to the full capacity of the instrument. Certified deadweights or master load cells should be used,
along with appropriate mounting brackets and fixtures. Caution should be taken while handling such
equipment.

10.2 Calibration Procedure

1. Select

Calibration from the menu. The display will appear as follows: