Checkline Series-5 User Manual
Page 19
Series 5 Digital Force Gauges
User’s Guide
18
Units
LB
Switch units to pound-force
OZ
Switch units to ounce-force
KG
Switch units to kilogram-force
G
Switch units to gram-force
N
Switch units to Newtons
MN
Switch units to Milli-Newtons
KN
Switch units to Kilo-Newtons
Basic Functions
CUR
Current mode (real time mode) for primary reading
PT
Peak Tension mode for primary reading
PC
Peak Compression mode for primary reading
CLR
Clear peaks
Z
Zero display and perform the CLR function
Filters
FLTCn
Digital filter for displayed readings
FLTPn
Digital filter for current readings
n= 0-10, filter = 2
n
, ex: n=0= no filter, n=10=1024 samples averaged
Memory & Statistics
MEM
Transmit all stored readings
STA
Transmit
statistics
Set
Points
SPHD
Disable high set point
SPLD
Disable low set point
SPHn
High set point. n=value (+ for compression, - for tension)
SPLn
Low set point. n=value (+ for compression, - for tension)
Note: High set point value must be greater than low set point value if both values
are set to the same polarity.
USB/RS-232
Communication
FULL
USB/RS-232 transmission with units
NUM
USB/RS-232 transmission without units (only numeric values)
AOUTn
Auto-transmit n times per second n=1,2,5,10,25,50,125,250. 0=disabled
Note: n = 1 = yields 50 times per second. This is provided for backward
compatibility with legacy gauges.
Mitutoyo
Communication
MIT
Enable Mitutoyo output
MITD
Disable Mitutoyo output
POL
Mitutoyo output with polarity (+ for compression, - for tension)
NPOL
Mitutoyo output without polarity (absolute value)
PM
Print/send data to a Mitutoyo-compatible device
Averaging
A
Enable Average mode
AD
Disable Average mode
AM
Select Average mode (if enabled) for primary reading
ATn
Average time. n=0.1-300.0 seconds
DELn
Initial delay. n=0.1-300.0 seconds
TRFn
Trigger force. n=value (+ for compression, - for tension)