beautypg.com

BUCHI Rotavapo R-220 SE User Manual

Page 82

background image

8 Troubleshooting

82

R-220 SE Operation Manual, Version F

Error

Addition error indicator on display

Possible cause

Remedy

E12 ‘FLOW’ message appears.

Cooling medium flow insufficient.

• Increase cooling water flow and

press the “Aerate” button to confirm
the error message.

• Adjust sensor.

• Check cable and plug of the flow

sensor.

• Exchange sensor if defective or run

evaporation without flow sensor.

Call service if problem is persistent.

E13 ‘LEVEL 1’ message appears. Filling level reached.

• Empty receiving vessel 1 and press

the “Aerate” button to confirm the
error message.

• Check cable and plug of the flow

sensor.

• Exchange sensor if defective or run

evaporation with user controlled
filling level.

Call service if problem is persistent.

E14 ‘LEVEL 2’ message appears. Filling level reached.

• Empty receiving vessel 2 and press

the “Aerate” button to confirm the
error message.

• Check cable and plug of the flow

sensor.

• Exchange sensor if defective or run

evaporation with user controlled
filling level.

Call service if problem is persistent.

E26 Heating bath icon flashes.

Bath temp. sensor shortcut.

• Restart the unit.

Call service if problem is persistent.

E27 Condenser icon flashes.

Vapor temp. sensor shortcut.

• Switch off the unit. Check sensor

cable and plug.

• Exchange sensor if defective or run

evaporation without vapor temp.
display.

Call service if problem is persistent.

E28 Condenser icon flashes.

Cooling water temp. sensor
shortcut.

• Switch off the unit. Check sensor

cable and plug.

• Exchange sensor if defective or run

evaporation without cooling medium
temp. display.

Call service if problem is persistent.