8 Troubleshooting
82
R-220 SE Operation Manual, Version F
Error
Addition error indicator on display
Possible cause
Remedy
E12 ‘FLOW’ message appears.
Cooling medium flow insufficient.
• Increase cooling water flow and
press the “Aerate” button to confirm
the error message.
• Adjust sensor.
• Check cable and plug of the flow
sensor.
• Exchange sensor if defective or run
evaporation without flow sensor.
Call service if problem is persistent.
E13 ‘LEVEL 1’ message appears. Filling level reached.
• Empty receiving vessel 1 and press
the “Aerate” button to confirm the
error message.
• Check cable and plug of the flow
sensor.
• Exchange sensor if defective or run
evaporation with user controlled
filling level.
Call service if problem is persistent.
E14 ‘LEVEL 2’ message appears. Filling level reached.
• Empty receiving vessel 2 and press
the “Aerate” button to confirm the
error message.
• Check cable and plug of the flow
sensor.
• Exchange sensor if defective or run
evaporation with user controlled
filling level.
Call service if problem is persistent.
E26 Heating bath icon flashes.
Bath temp. sensor shortcut.
• Restart the unit.
Call service if problem is persistent.
E27 Condenser icon flashes.
Vapor temp. sensor shortcut.
• Switch off the unit. Check sensor
cable and plug.
• Exchange sensor if defective or run
evaporation without vapor temp.
display.
Call service if problem is persistent.
E28 Condenser icon flashes.
Cooling water temp. sensor
shortcut.
• Switch off the unit. Check sensor
cable and plug.
• Exchange sensor if defective or run
evaporation without cooling medium
temp. display.
Call service if problem is persistent.