Ab c – BUCHI Rotavapo R-220 SE User Manual

Page 65

background image

5 Putting into operation

65

R-220 SE Operation Manual, Version F

After readjustment, perform a vacuum test-run without heating bath and sample medium.

NOTE
Perform the adjustment in single-turn steps only. Inbetween, check for proper tension with the empty
evaporating flask.

5 .19

Installation and usage of a shut-off tap

The shut-off tap provides a three stage valve mechanism with a spring-loaded seal stamp. This
prestressing assures safe seal under normal conditions.

Turn the lower, white part clockwise to install the tap at the glassware thread. Mount it handtight only!

a

b

c

“OFF” stage

“Medium throughput” stage

“Maximum throughput” stage

Installation

a

In this position, the shut-off tap completely blocks the flow of the medium.

b

Starting with the tap in “OFF” stage – turn the upper, black part clockwise to reach the
“Medium throughput” stage. A noticeable drop in resistance can be felt between every stage.

c

Starting with the tap in “Medium throughput” stage – turn the upper, black part clockwise to
reach the “Maximum throughput” stage.