beautypg.com

3 .2 technical data – BUCHI Rotavapo R-220 SE User Manual

Page 18

background image

3 Technical data

18

R-220 SE Operation Manual, Version F

3 .2

Technical data

Technical data Rotavapor R-220 SE

Electrical connection data, single phase

Voltage

220 – 240 V

± 10% (1P, N, G)

Frequency

50/60 Hz

Max. power consumption

3800 W (with 2×1800 W heating elements)

Overvoltage category

II

Built-in mains supply socket, switched

220 – 240 VAC, 50/60 Hz, 2 A

Electrical connection data, three phase

Voltage

400 V

±10% (3P, N, G)

Frequency

50/60 Hz

Max. power consumption

3800 W or 4400 W (with 2×2100 W heating elements)
6500 W with 6,3 kW heating (with 3x2100 W heating elements)

Overvoltage category

II

Built-in mains supply socket, switched

220 – 240 VAC, 50/60 Hz, 2 A

Electrical connection data, Japan version

Voltage

200 V

± 10% (1P, N, G)

Frequency

50/60 Hz

Max. power consumption

3800 W (with 2×1800 W heating element)

Overvoltage category

II

Built-in mains supply socket, switched

220 – 240 VAC, 50/60 Hz, 2 A

Environmental conditions

Pollution degree

2 (for indoor use only)

Temperature

5 – 40 °C

Altitude (above sea level)

up to 2000 m

Humidity (curve parameter)

Maximum relative humidity 80% up to 31 °C, then decreasing
linearly to 50% relative humidity at 40 °C

Dimensions and weight

Height

1430 mm – 1800 mm, depending on glass assembly

Width

1200 mm

Depth

710 mm max., depending on glass assembly

Bath diameter (inner)

430 mm

Weight

Approx. 70 kg without glass

Display

Bath temperature

(1 °C steps)

Cooling water temperature (optional)

(1 °C steps)

Vapor temperature

(1 °C steps)

Setting for the rotational speed

(1 rpm steps)

Setting for the bath temperature

(1 °C steps)

Actual vacuum

(1 mbar steps)

Setting for the vacuum

(1 mbar steps)

Cooling

Cooling water flow rate

120–200 liters/hour (variable)

Limitations

Max. 2.7 bar absolute pulse-free