3 .2 technical data – BUCHI Rotavapo R-220 SE User Manual
Page 18
![background image](/manuals/656056/18/background.png)
3 Technical data
18
R-220 SE Operation Manual, Version F
3 .2
Technical data
Technical data Rotavapor R-220 SE
Electrical connection data, single phase
Voltage
220 – 240 V
± 10% (1P, N, G)
Frequency
50/60 Hz
Max. power consumption
3800 W (with 2×1800 W heating elements)
Overvoltage category
II
Built-in mains supply socket, switched
220 – 240 VAC, 50/60 Hz, 2 A
Electrical connection data, three phase
Voltage
400 V
±10% (3P, N, G)
Frequency
50/60 Hz
Max. power consumption
3800 W or 4400 W (with 2×2100 W heating elements)
6500 W with 6,3 kW heating (with 3x2100 W heating elements)
Overvoltage category
II
Built-in mains supply socket, switched
220 – 240 VAC, 50/60 Hz, 2 A
Electrical connection data, Japan version
Voltage
200 V
± 10% (1P, N, G)
Frequency
50/60 Hz
Max. power consumption
3800 W (with 2×1800 W heating element)
Overvoltage category
II
Built-in mains supply socket, switched
220 – 240 VAC, 50/60 Hz, 2 A
Environmental conditions
Pollution degree
2 (for indoor use only)
Temperature
5 – 40 °C
Altitude (above sea level)
up to 2000 m
Humidity (curve parameter)
Maximum relative humidity 80% up to 31 °C, then decreasing
linearly to 50% relative humidity at 40 °C
Dimensions and weight
Height
1430 mm – 1800 mm, depending on glass assembly
Width
1200 mm
Depth
710 mm max., depending on glass assembly
Bath diameter (inner)
430 mm
Weight
Approx. 70 kg without glass
Display
Bath temperature
(1 °C steps)
Cooling water temperature (optional)
(1 °C steps)
Vapor temperature
(1 °C steps)
Setting for the rotational speed
(1 rpm steps)
Setting for the bath temperature
(1 °C steps)
Actual vacuum
(1 mbar steps)
Setting for the vacuum
(1 mbar steps)
Cooling
Cooling water flow rate
120–200 liters/hour (variable)
Limitations
Max. 2.7 bar absolute pulse-free