5 Putting into operation
49
R-220 SE Operation Manual, Version F
Test results
Contaminant
Concentration germs/ml Sample 1 germs/ml
Sample 2 germs/ml
Sample 3 germs/ml
Escherichia coli ATCC 25922
10
3
0
0
0
Escherichia coli ATCC 25922
10
4
0
0
0
Escherichia coli ATCC 25922
10
5
0
0
0
Escherichia coli ATCC 25922
10
6
0
0
0
Staphylococcus aureus ATCC 25923
10
3
0
0
0
Staphylococcus aureus ATCC 25923
10
4
0
0
0
Staphylococcus aureus ATCC 25923
10
5
0
0
0
Staphylococcus aureus ATCC 25923
10
6
0
0
0
Saccharomyces cerevisiae NCYC 79
10
3
0
0
0
Saccharomyces cerevisiae NCYC 79
10
4
0
0
0
Saccharomyces cerevisiae NCYC 79
10
5
0
0
0
Saccharomyces cerevisiae NCYC 79
10
6
0
0
0
Geobac. stearothermophilus ATCC 7953
10
3
40
0
0
Geobac. stearothermophilus ATCC 7953
10
4
560
-
-
Geobac. stearothermophilus ATCC 7953
10
5
3600
0
0
Geobac. stearothermophilus ATCC 7953
10
6
46000
0
0
5 .12
General installation instructions for glassware
Small tolerances of glassware parts are conditional of manufacturing. Thus, it is necessary to indi-
vidually adjust fixtures and holders at installation to avoid stress at the components.
It is recommended to install a glassware setup starting with the distribution piece at the gearbox. All
other parts will subsequently be aligned towards the Rotavapor via the distribution piece. The drawing
below shows an example installation of a complex setup. For other setups proceed adaptive (e.g.
leave parts / steps out).