beautypg.com

BUCHI Rotavapo R-220 SE User Manual

Page 19

background image

3 Technical data

19

R-220 SE Operation Manual, Version F

Technical data Rotavapor R-220 SE

Heating bath

Heating medium

Depending on heating wattage

3600 W (water or other dedicated heating transfer media)
4200 W (water only!)
6300 W (water only!)

Bath capacity

20 l, without flask immersed

Distillation capacity (ethanol)

Up to 13.5 liters/hour 3600 W heater
Up to 16 liters/hour 4200 W heater
Up to 18 liters/hour 6300 W heater

Temperature range

Room temperature up to 180 °C

Heating power

230 V 1-phase; 3600 W (3 W/cm

2

)

optional 4200 W (3,5 W/cm

2

, for water only)

Regulation precision

+/- 2 °C

Bath over temperature protection

• Independent electronic monitoring circuit, mechanical

reset

• Error if the actual temperature is 15 °C above the setpoint

Bath lift

Linear drive, safety class

Battery

Lead-acid, 12 V

Rotational drive

Motor

200 V; 1-phase; 0.6 A at 50 Hz

Control

Electronic, with softstart

Rotation speed

5 – 140 rpm

Accuracy

+/- 1 rpm at 5 rpm to +/- 5 rpm at 150 rpm

Behavior in case of rotation error

• Bath lowered, heater off, rotation off

• Indication of the error

• Reset possible

Glass assembly

Pressure in the glass assembly

0 mbar up to ambient pressure

Tightness

< 1 mbar/min

Required external vacuum pump performance
for vacuum evaporation

2 – 4 m

3

/h

Sensors

Vapor temperature

PT-1000, 2-wire

Bath temperature

PT-1000, 2-wire

Bath level sensor

n.a.

Pressure

Ceramic, capacitive