5 Putting into operation

50

R-220 SE Operation Manual, Version F

a

b

c

d

e

a

Start with the distribution piece and mount it at the gearbox.

b

Proceed with the expansion vessel. Mind upright installation!

c

• Use the u-shaped bridge tube to roughly align the first condenser to the expansion

vessel.

• Fix the u-tube at the expansion vessel and subsequently bring in the condenser.
• Fasten the condenser. Fix the tube to the condenser.

d

• Use the y-shaped bridge tube to roughly align the outlet of the second condenser to the

first condenser from below.

• Fix the y-piece at the first condenser outlet.
• Subsequently bring in the second condenser, fasten it and fix the y-piece.

e

Align the receiving flask to the y-piece and fasten the clamp.