5 Putting into operation

56

R-220 SE Operation Manual, Version F

5 .15

Heating bath setup for 2×1800 W heating element

The heating element holder is labelled with the heating power data. The heating bath of the 2×1800 W
version can be filled with different kinds of liquid heating transfer medium. Depending on the
maximum heating bath temperature for the target application, a suitable medium must be used.

How to set the maximum setpoint temperature
Limiting the maximal temperature set-point to a value below 180 °C can help to eliminate processing
errors, e.g. when working with heat-sensitive substances.

AERATE

Temp

+

Set Temp

UP

DOWN

UP

a

b

Store setting

Keep the "TEMP UP" button pressed while switching on
the Rotavapor to enter the setup menu.

Set the temperature limit via the "TEMP UP" and "TEMP
DOWN" buttons. The set value is displayed as bath
temperature value on the display.

➥ Press "AERATE" button to store the setting and to

leave the menu.