5 Putting into operation
58
R-220 SE Operation Manual, Version F
5 .15 .1
Water as heating transfer medium
With the heating bath in lowest position, install the level drain output tube with a level difference of
> 5 cm towards the sink to ensure safe draining. Fixate all tube connections with hose clamps!
NOTICE
Risk of damage caused by heating transfer mediums.
• Do not fill the heating bath to overflow
• Do not leave the system unattended when replenishment is active
• Check the level drain for proper drainage function in lowest heating bath position
Do NOT connect the replenishment tap to a water source in case
a heating transfer medium different to water (e.g. glycerin) is used!
!
!