- 49 -

2. Spindle jammed below zero

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for

repair

Display Reading Over 100

1. Overrange “EEE” (in %, cP, and SS)

• Change spindle and/or speed

Viscometer Will Not Return to Zero

1. Viscometer is not level

• Check with spindle out of the sample
• Adjust the laboratory stand

2. Pivot point or jewel bearing faulty

• Perform calibration check
• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for

repair

Display Reading Will Not Stabilize

1. Check for erratic spindle rotation

• Verify power supply
• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for

repair

2. Bent spindle or spindle coupling

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for

repair

3. Temperature fluctuation in sample fluid

• Use temperature bath control

4. Special characteristic of sample fluid

• Refer to Appendix C