Brookfield DV-II+ Digital Viscometer User Manual
Page 24
- 24 -
Each of the five available cones has a specific sample volume, as shown in Table A1. Note that the
correct amount of sample fluid should cover the cone face and back up over the edge less than 1mm as
shown in (Figure A8).
Figure A8
Notes:
a) The cup may be removed without resetting the gap.
b) Remove the spindle from the viscometer when you clean it.
c) Reset the hit point every time the spindle is attached.
Table A1
Cone
Sample Volume
Cone Angle
CP-40
0.5 ml
0.8
°
CP-41
2.0 ml
3.0
°
CP-42
1.0 ml
1.565
°
CP-51
0.5 ml
1.565
°
CP-52
0.5 ml
3.0
°
Sample
Cup
Cone
Less than
1 mm
Cone
Less Than
1 mm
Cup
Sample
Calibration Procedure for Cone/Plate Viscometers:
1) Follow the procedures outlined in Appendix A for mechanically adjusting the setting of the cone
to the plate.
2) Refer to Table A1, and determine the correct sample volume required for the spindle to be
utilized.
3) Select a viscosity standard fluid that will give viscosity readings between 10% and 100% of full
scale range. Refer to Appendix B for viscosity ranges of cone spindles. Consult with Brookfield
or an authorized dealer to determine which fluid is appropriate.