Negative % (Torque) will be displayed as shown in Figure 17:

Figure 17

cP (Viscosity) and SS (Shear Stress) values will be displayed as "- - - -" when the % (Torque) is
below zero.

Figure 18 is an example of the printed output of each of the above conditions.

RPM=50 MD=RV SPDL=29 %=51.4 cP=10280 D/CM2=1285 1/SEC=12.5 T=25.5C

Normal Print String

RPM=50 MD=RV SPDL=29 %=EEEE cP=EEEE D/CM2=EEEE 1/SEC=12.5 T=25.5C

Over-Range

?RPM=50 MD=RV SPDL=29 %=5.2 cP=1040 D/CM2=130 1/SEC=12.5 T=25.5C

Under-Range

RPM=50 MD=RV SPDL=29 %=-0.1 cP=---- D/CM2=---- 1/SEC=0.0 T=25.5C

Negative Torque

Figure 18

II.7

Temperature Display

The DV-II

+

will display the temperature measured by its RTD temperature probe. Temperature may

be displayed in either ˚C (Centigrade) or ˚F (Fahrenheit) units depending upon the start-up
operations. If the DV-II

+

is started normally, the default temperature display will be in ˚C

(Centigrade) units as shown in the Figure 19:

Figure 19

However, if you press and hold the SELECT DISPLAY key while turning on the DV-II

+

the

temperature will be displayed in ˚F (Fahrenheit).

If you turn on the DV-II

+

with the temperature probe disconnected, or remove the temperature probe

at any point after power-up, the display will indicate “- - - -

°

C”.

The four “dashes” indicate the absence of the probe. If you were displaying temperature in
Fahrenheit units the C would be displayed as an F.