beautypg.com

Brookfield DV-II+ Digital Viscometer User Manual

Page 33

background image

- 33 -

The full scale viscosity range for any DV-II+ model and spindle may be calculated using the

equation:

Where:

TK

= DV-II+ Torque Constant from Table D2

SMC = Spindle Multiplier Constant from Table D1

The Shear Rate calculation is:

Shear Rate (1/Sec) = SRC * RPM

Where:

SRC = Shear Rate Constant from Table D1

Full Scale Viscosity Range [cP] = TK SMC

10,000
RPM

*

*