Please check to be sure that you have received all components, and that there is no damage. If you
are missing any parts, please notify Brookfield Engineering or your local Brookfield agent immedi-
ately. Any shipping damage must be reported to the carrier.

I.2

Utilities

Input Voltage:

115 VAC or 230 VAC

Input Frequency:

50/60 Hz

Power Consumption:

Less than 20 WATTS

Power Cord Color Code:

United States

Outside United States

Hot (live)

Black

Brown

Neutral

White

Blue

Ground (earth)

Green

Green/Yellow

I.3

Specifications

Speeds:

LVDV-II

+

:

0.0, 0.3, 0.6, 1.5, 3, 6, 12, 30, 60, 0.0,
0.5, 1, 2, 2.5, 4, 5, 10, 20, 50, 100

RV/HA/HBDV-II

+

:

0.0, 0.5, 1, 2, 2.5, 4, 5, 10, 20, 50, 100,
0.0, 0.3, 0.6, 1.5, 3, 6, 12, 30, 60

Weight:

Gross Weight

20

lbs.

9 kg.

Net Weight

17

lbs.

7.7 kg.

Carton Volume

1.65 cu. ft.

0.05 m

3

Temperature sensing range:

-100˚C to 300˚C (-148˚F to 572˚F)

Analog Torque Output:

0 - 1 Volt DC (0 - 100% Torque)

Analog Temperature Output:

0 - 4 Volts DC (10mv / ˚C)

RS232 Compatible Serial Port for use with an attached printer or PC.

I.4

Set-Up

1)

Place the upright rod into the base (refer to assembly instructions in Appendix H). The rack
gear and clamp assembly should face the front of the base. The upright rod is held in place
with the jam nut which is attached from the bottom of the base. Tighten this nut with a
suitable wrench (spanner). Attach leveling feet.

2)

Insert the mounting handle on the back of the DV-II

+

Viscometer into the hole on the clamp

assembly. Be sure that the clamp screw, VS-24Y, is loose.

3)

Tighten the VS-24Y clamp screw. Adjust viscometer to be as close to level as possible while
tightening VS-24Y.

4)

Connect the RTD probe to the socket on the rear panel of the DV-II

+

.

5)

The Viscometer must be leveled. The level is adjusted using the three leveling screws on
the base. Adjust so that the bubble level on top of the DV-II

+

is centered within the circle.

Note: Check level periodically during use.