Creating a New Calibration using Two Points
1.
From the EasyData main screen, select Setup, and
then select CH1: (or the appropriate channel).
2. Select
from the Sensor Setup screen.
3. Select
. A menu is displayed showing the different
calibration methods.
4. Select
Two Point Live….
5.
Place your sensor into the necessary condition for the
first calibration point (e.g. a force sensor with no mass
hanging from it or a pH sensor in a pH buffer). Monitor
the displayed Potential (V). When the value is stable,
select
.
6.
Enter the known value for the first calibration condition
(e.g. “0” for a force of zero N), and select
.
7.
Place your sensor into the necessary condition for the
second calibration point (e.g. a force sensor with a 200
g mass hanging from it). Monitor the displayed Potential
(V). When the value is stable, select
.