56

Troubleshooting and maintenance

Welch Allyn AcuityLink Clinician Notifier

Maintenance

For further maintenance instructions, refer to the directions for use provided by the
mobile device manufacturer.

Mobile device battery

Main battery

To recharge the battery

1.

If you do not see a Clinician Notifier program window on your mobile device screen,

tap

, and then tap

Clinician Notifier.

A Clinician Notifier screen appears.

2.

Tap

in the upper right corner of any page.

A confirmation popup window appears.

3.

Tap YES.

You are logged to log out of the Clinician Notifier program.

4.

Recharge the device as specified by the device manufacturer.

To replace the battery

Replace the battery when it no longer holds a charge.

1.

Turn off the device.

2.

Remove the battery.

3.

Insert a fresh battery according to the device manufacturer’s instructions.

4.

Recycle the battery according to your local protocol.

WARNING Connecting mobile devices to the Internet, or running too many
programs in the foreground or background while the Clinician Notifier program is
running, can greatly reduce the battery life and memory of the devices, thereby
compromising alarm delivery.

WARNING Always recharge a mobile device when it is not in use and at first
notification of low batteries. A device with low batteries might not indicate or
appropriately clear all of a patient’s alarms. Allowing batteries to fully discharge
can result in depletion of the backup battery and loss of Clinician Notifier
configurations.