Respond to alarms and alerts
Welch Allyn AcuityLink Clinician Notifier
Respond to patient alarms
There are two types of patient alarm popup windows:
•
Single alarm popup window
•
Multiple alarms popup window
Single alarm
If only one alarm is active on your mobile device, this Alarm popup window appears in the
foreground of your screen.
If more than one alarm occurs at once for the same patient, the patient’s highest priority
alarm is indicated in the popup window.
A single alarm popup window is always associated with a tone, vibration or both.
To temporarily stop alarm notification
•
Tap Suspend to suspend the alarm.
The Suspend button changes to Resume. If you want to resume the alarm, tap
Resume.
Notification of all alarms for this patient is suspended for 90 seconds on all mobile
devices that received the alarm. However, new lethal arrhythmia alarms and
apnea alarms for the patient interrupt the suspend state.
Tones are also suspended for this patient at the Acuity Central Station and the
patient monitor.
•
Tap Busy to indicate that you are busy.
Notification for this alarm stops on your device only.
If someone suspends the alarm and it remains active after the 90 second period,
notification returns to your device.
Event Start time indicates the time that the
notification occurs. If the event re-occurs after
suspension, Event Start time indicates the re-
occurrence time.