beautypg.com

Calibration temperature – Torbal AGCN200 User Manual

Page 16

background image

Instruction Manual - TORBAL AGCN Series

3. Press the T Key again when “CAL TM” is displayed.

4. Time interval settings will begin to display sequentially. The current time interval setting will be marked with

“o” located in the lower left corner of the display.

5. Automatic calibration can be set to calibrate between 1, 2, 3, 4, 5, and 6 hours (i.e., if set to 3, calibration will

be performed at least once every three hours).

6. To select a time interval, press the T Key when the desired setting is displayed.

To disable the time interval calibration, press the T Key when OFF is displayed.

8.1.4.Calibration Temperature

Internal calibration can be configured to calibrate automatically within a change of temperature. To set a calibration
temperature interval, follow these steps:

1. Enter the main menu by pressing the MENU key.

2. Enter the calibration set-up by pressing the T Key when “CALIBR” is displayed.

16

This manual is related to the following products: