Torbal AGCN200 User Manual
Page 40

Instruction Manual - TORBAL AGCN Series
7. To key in the volume of the plunger, use the following keys: the C key to increment a digit, the T key to
accept and go to the next digit, the P key to insert a decimal, and MENU to accept the setting.
8. Once the volume of the plunger has been entered, command “PRESS TARE” will be displayed indicating to
tare the scale.
9. Once the scale has been tared, hang the plunger on the wire support frame of the density apparatus.
10. The weight of the plunger will be displayed. Once the weight is stable, press the P key. Command “M1” will
be displayed, indicating that the weight of the plunger has been stored in the formula.
40