Calibration time – Torbal AGCN200 User Manual
Page 15
Instruction Manual - TORBAL AGCN Series
9. Once the new calibration value is stored, the command “UNLOAD” will be displayed, indicating that this
calibration weight should be removed from the pan.
10. Once the weight is removed, the scale will continue with calibration adjustment. “WAIT” will be displayed
during the remainder of the adjustment process. Once finished, the scale will automatically return to the
weighing mode.
8.1.3.Calibration Time
Internal calibration can be configured to calibrate automatically after a set interval of time. To set a calibration time
interval, follow these steps:
1. Enter the main menu by pressing the MENU key.
2. Enter calibration set-up by pressing the T Key when “CALIBR” is displayed.
15