Character Description
C
Begin span calibration
xD
Set 1 second print delay (set x = 0 for OFF, or x = 1 for ON)
PM
Application mode.
xFL
Set Averaging Filter Level 1= Low, 2= Medium, 3= High
L
Begin linearity calibration
P
Print displayed weight (stable or unstable)
T
Same as pressing Tare key.
PV
Version: print name, software revision and LFT ON (if LFT is set ON)
xAL
Set Auto-Zero level to x. x = 0 for OFF, x = 1 for 0.5d, x=2 for 1.0d, x=3 for 2.0d, x=4 for
5.0d.
Esc R
Resets Setup and Print menus to factory defaults
On
Turns balance on
Off
Turns balance off
x%
Set % reference weight (x) in current unit
x#
Set PC reference weight (x) in current unit
P#
Print PC reference weight
P%
Print percent reference weight
xM
Set current Application mode to x. x = 1 for Weighing, x = 2 for Parts Counting, x = 3 for
Animal Weighing, x = 4 for Percent Weighing, x = 5 for Check Weighing, x = 6 for G/N/T,
x = 7 for Filling, x = 8 for Differential, x = 9 for Formulation, x = 10 for Statistical, x = 11
for SQC, 12 = for Density
xAW
Set Animal Weigh Level to x. x = 1 for LOW, x = 2 for MEDIUM, x = 3 for HIGH.
xAM
Set animal mode. Where x is 1=Manual, 2=Semi and 3=Auto
SAW
Start Animal cycle.
xT
Download Tare value in current unit. Sending 0T clears tare.
PID
Print current user ID string
xID
Program user ID string. 1-8 characters
AC
Abort calibration
xUC
Set user defined weight
UC
Initiate user calibration
IC
Initiate internal calibration
PTIME
Print current time
mm/dd/yyDATE
Set date
hh:mm:ssTIME
Set time
PDATE
Print current date
5. TECHNICAL DATA
5.1 RS232 Commands
Field: Weight Unit Stab CR LF
Length: Max 9
5 1 1 1