Menu, Power on/off, Stabilization – Rice Lake Ohaus Explorer Pro High Capacity Instuction Manual User Manual
Page 15: 4 applications, 1 weighing

Explorer
Pro
EN-11
3.4 Applications
The Explorer Pro balance contains Weighing, Parts Counting,
Animal Weighing, Percent Weighing, Check Weighing, G/N/T,
Statistics, SQC, Formulation, Differential Weighing and Filling
applications. Before using any of these applications, they
must be turned on first before they can be accessed. A
choice can be made to turn on or off as many of these
applications as required. The balance default setting has the
weighing mode turned on and all other modes are off.
To select applications to be turned on or off, press the Menu
button and using the arrow buttons, scroll to Application Modes,
then press the Enter button.
MENU
Calibration
Balance Options
Readout
Application Modes
Units
Interface
Print Options
GLP Print Options
Library
Lockout
Factory Reset
Exit
APPLICATION MODES
Weighing
:
On
Parts Counting
: Off
Animal Weighing
: Off
Percent Weighing
: Off
Check Weighing
: Off
G/N/T
: Off
Filling
: Off
Differential
: Off
Formulation
: Off
Statistics
: Off
SQC
: Off
Density
: Off
By pressing the Enter button and using the arrow buttons, each
item on the Application Modes screen can be turned on or off.
When finished, press the Menu button, the balance returns to
weighing mode. All settings are saved.
To use any of the modes that have been turned on, press the
Mode button repeatedly to cycle through all of the application
modes.
3.4.1 Weighing
The Explorer Pro balance is shipped with grams only enabled.
The balance can be used immediately after calibration has
been performed. When the balance is to be used with other
units of measure, the desired unit must be enabled.
USE
Zero the balance. Place objects or material to be weighed on the
pan. Example indicates a 200 gram weight on a 6100g balance.
Wait for the stability indicator to appear before reading the weight.
WEIGHING
01/01/03 03:05:45: PM
UNITS
2nd UNITS
CALIBRATE
200.00
g
*
0.00g
0g
6100g
WEIGHING WITH A CONTAINER
With no load on the pan, zero the balance.
Place an empty container on the pan. Its weight is displayed.
Tare the balance. The container’s weight is stored in memory.
Add material to the container. As material is added, its net weight
is displayed.
Removing the container and material from the pan will cause the
balance to display the container’s weight as a negative number.
The tared container weight will remain in memory until the balance
is zeroed again or the balance is turned off.