Addendum 1: quick setup and configuration, 1 configure the sbi-521 to a floor scale, 1 setting the capacity and count by – Salter Brecknell WB-521 Series User Manual
Page 75: Configure the sbi-521 to a floor scale, Setting the capacity and count by
SBI521 Service Manual
73
Addendum 1: Quick Setup and Configuration
9.1 Configure the SBI-521 to a Floor Scale
9.1.1 Setting the Capacity and Count by
1.
Refer to the next page for wiring and of sense line jumpers placement.
2.
Press the [HOLD/SETUP] key for 3 seconds. The indicator shows Config to
indicate that you are in Setup Menu mode.
3.
Press the [TARE] key once. (Tare is also used as an Enter key).
4.
The display will show the Audit Trail counter then cycle to
CFG.ON.
4a.
If the display shows
CFG.OFF, then remove the CAL seal on the back of the
indicator. Move the switch to the opposite position. Cycle power on the indicator
and go back to step 1.
5.
Press the [TARE] key once. Display will show
RESET.
6.
Press the [PRINT] key until the display shows
PRIM.N.
Table 9.1 Reference to Number of Divisions and Division Size
7.
Press the [TARE] key once. The display will show
003000 (this is the
number of divisions).
8.
Press the [UNIT] key 3 times and the
3 should be flashing.
9.
Press the [N/G] key 2 times until the
5 is displayed. (example for 5000 x 1 lb)
10.
Press the [TARE] key once.
11.
Press the [PRINT] key once until the display shows
PRIM.D.
12.
Press the [TARE] key once. Make sure the display shows
1.
13.
Press the [TARE] key once. The display should show
PRIMM.D.
14.
Press the [PRINT] key once until the display shows
PRIM.UT.
15.
Press the [TARE] key once. The display should show
KG.
16.
Press the [PRINT] key once.
17.
Press the [TARE] key once.
18.
Press the red [ZERO] key 2 times.
Capacity
Required
PRIM.N
PRIM.D
2000
4000
.5
5000
5000
1
10000
5000
2