SBI521 Service Manual

73

Addendum 1: Quick Setup and Configuration

9.1 Configure the SBI-521 to a Floor Scale

9.1.1 Setting the Capacity and Count by

1.

Refer to the next page for wiring and of sense line jumpers placement.

2.

Press the [HOLD/SETUP] key for 3 seconds. The indicator shows Config to
indicate that you are in Setup Menu mode.

3.

Press the [TARE] key once. (Tare is also used as an Enter key).

4.

The display will show the Audit Trail counter then cycle to

CFG.ON.

4a.

If the display shows

CFG.OFF, then remove the CAL seal on the back of the

indicator. Move the switch to the opposite position. Cycle power on the indicator
and go back to step 1.

5.

Press the [TARE] key once. Display will show

RESET.

6.

Press the [PRINT] key until the display shows

PRIM.N.

Table 9.1 Reference to Number of Divisions and Division Size

7.

Press the [TARE] key once. The display will show

003000 (this is the

number of divisions).

8.

Press the [UNIT] key 3 times and the

3 should be flashing.

9.

Press the [N/G] key 2 times until the

5 is displayed. (example for 5000 x 1 lb)

10.

Press the [TARE] key once.

11.

Press the [PRINT] key once until the display shows

PRIM.D.

12.

Press the [TARE] key once. Make sure the display shows

1.

13.

Press the [TARE] key once. The display should show

PRIMM.D.

14.

Press the [PRINT] key once until the display shows

PRIM.UT.

15.

Press the [TARE] key once. The display should show

KG.

16.

Press the [PRINT] key once.

17.

Press the [TARE] key once.

18.

Press the red [ZERO] key 2 times.

Capacity

Required

PRIM.N

PRIM.D

2000

4000

.5

5000

5000

1

10000

5000

2