SBI521 Service Manual

41

Table 4.2 CONFIG Menu Choices and Explanations

CONFIG

SubMenu1

SubMenu2

Option

Default

Parameter Description

Comment

CFG.ON

CFG.OFF

Seal switch position

The display will show whether the
seal switch is in the ON or Off
position. This parameter can’t be
changed within the software.

RESET NO

NO

Reset Config menu
parameters to default
setting

YES

REGULA

none

USA

Select the standard in
which the scale will
comply: USA, Canada,
Europe

*None = not legal for trade.

USA

Canada
Europe

PRIM.N

100 -

100,000

3000

Primary full scale value

*If (REGULA) ≠ none, the max is
10,000

PRIM.D

0.0001

1

The division value under
primary unit

The division value under second
unit is automatically determined by
the indicator according to the
division value under primary unit.

0.0002
0.0005

0.001
0.002
0.005

0.01
0.02
0.05

0.1
0.2
0.5

1
2
5

10
20
50

PRIM.Ut

kg

kg

Primary unit

Select the primary unit from kg or lb.
The second unit is lb if kg is
selected as the primary unit.
*The calibration standard weight
must be in primary unit!

lb

SECND.N

100-

125,000

3000

Second scale full scale
value

The division number under second
unit. The maximum is
1.25*(PRIM.N).
*If (REGULA)≠none, the max is
10,000