Salter Brecknell WB-521 Series User Manual
Page 43
SBI521 Service Manual
41
Table 4.2 CONFIG Menu Choices and Explanations
CONFIG
SubMenu1
SubMenu2
Option
Default
Parameter Description
Comment
CFG.ON
CFG.OFF
Seal switch position
The display will show whether the
seal switch is in the ON or Off
position. This parameter can’t be
changed within the software.
RESET NO
NO
Reset Config menu
parameters to default
setting
YES
REGULA
none
USA
Select the standard in
which the scale will
comply: USA, Canada,
Europe
*None = not legal for trade.
USA
Canada
Europe
PRIM.N
100 -
100,000
3000
Primary full scale value
*If (REGULA) ≠ none, the max is
10,000
PRIM.D
0.0001
1
The division value under
primary unit
The division value under second
unit is automatically determined by
the indicator according to the
division value under primary unit.
0.0002
0.0005
0.001
0.002
0.005
0.01
0.02
0.05
0.1
0.2
0.5
1
2
5
10
20
50
PRIM.Ut
kg
kg
Primary unit
Select the primary unit from kg or lb.
The second unit is lb if kg is
selected as the primary unit.
*The calibration standard weight
must be in primary unit!
lb
SECND.N
100-
125,000
3000
Second scale full scale
value
The division number under second
unit. The maximum is
1.25*(PRIM.N).
*If (REGULA)≠none, the max is