Page 71: Audit trail parameters

background image

SBI521 Service Manual

69

7.2 Audit Trail Parameters

Two separate incrementing, non-resetable audit trail parameters are used by the Model
SBI-521 to indicate changes to various parameters or calibration.

7.2.1 View Configuration Counter

1.

Press and hold the [SETUP] key for 3 seconds.

CONFIG is displayed

2.

Press the [TARE] key.

2a.

The display will briefly show the number of times the configuration has been
changed and saved up to 999.

To Exit:

1.

Press the [ON/OFF] key.

Exit? is displayed.

2.

Press the [TARE] key to return to weigh mode.

7.2.2 View Calibration Counter

1.

Press and hold the [SETUP] key for 3 seconds.

CONFIG is displayed.

2.

Press the [FUNC] key twice.

Cal is displayed.

3.

Press the [TARE] key.

3a.

The display will briefly show the number of times the configuration has been
changed and saved up to 999.

To Exit:

1.

Press the [ON/OFF] key.

Exit? is displayed.

2.

Press the [TARE] key to return to the weigh mode.