2 audit trail parameters, 1 view configuration counter, 2 view calibration counter – Salter Brecknell WB-521 Series User Manual
Page 71: Audit trail parameters
SBI521 Service Manual
69
7.2 Audit Trail Parameters
Two separate incrementing, non-resetable audit trail parameters are used by the Model
SBI-521 to indicate changes to various parameters or calibration.
7.2.1 View Configuration Counter
1.
Press and hold the [SETUP] key for 3 seconds.
CONFIG is displayed
2.
Press the [TARE] key.
2a.
The display will briefly show the number of times the configuration has been
changed and saved up to 999.
To Exit:
1.
Press the [ON/OFF] key.
Exit? is displayed.
2.
Press the [TARE] key to return to weigh mode.
7.2.2 View Calibration Counter
1.
Press and hold the [SETUP] key for 3 seconds.
CONFIG is displayed.
2.
Press the [FUNC] key twice.
Cal is displayed.
3.
Press the [TARE] key.
3a.
The display will briefly show the number of times the configuration has been
changed and saved up to 999.
To Exit:
1.
Press the [ON/OFF] key.
Exit? is displayed.
2.
Press the [TARE] key to return to the weigh mode.