Salter Brecknell WB-521 Series User Manual
Page 44
42
SBI521 Service Manual
10N.DSP
NO
NO
Display weight at 10
times division number
under primary unit
*If (REGULA)= none this
parameter will not be available. T
YES
UNITS
0-6
3
Units key
0=kg
1=lb
2=lb:oz
3= kg,lb
4=kg,lb:oz
5=lb,lb:oz
6=kg,lb,lb:oz
MOTION
1-255
4
Motion window
1-255 = ±0.25d *(1-255)
*If (REGULA)=none, the max is 12
OVER.LD
0-100
0
Overload display
limitation
0=FS+9d
1-100=101%FS -200%FS.
*If (REGULA)=none, the max is 10
AD.FROM
ADC
ADC
Data of A/D comes from
ADC=local A/D chip on PCB.
COM2=COM2 interface.
NOTE: In legal trade application,
COM2 may be not allowed, Please
contact your local legal
metrology organization!
COM2
AD.H.SPD
NO
NO
Speed of the A/D
converter
no = 10Hz
yes = 80Hz
*If AD.FROM = COM2, this
parameter will not be displayed.
YES
ZRO.PNT
IZSM
0-100
10
Initial zero (power on
zero) point range
0=no limitation
1-100= (calibration zero point)
±1%FS - (calibration zero point)
±100%FS
*If (REGULA)=none, the max is 10
IN.IZSM
WEIGHT
WEIGHT Choose which weight as
current initial zero point
when current weight is in
IZSM range
WEIGHT= current weight
CAL.ZRO
CAL.ZRO= calibration zero
LAST.Z.T
LAST.Z.T=switch-off zero and tare
*If (REGULA) ≠none, the value is
fixed on WEIGHT
OV.IZSM
DSP.OVR
DSP.OVR Choose which weight as
current initial zero point
when current weight is
over IZSM range
DSP.OVR=display initial zero is over
WEIGHT
WEIGHT= current weight
CAL.ZRO
CAL.ZRO= calibration zero
LAST.Z.T
LAST.Z.T=switch-off zero and tare If
(REGULAR) ≠none, the value is
fixed on DSP.OVR
SAZSM
0-100
2
Zero key range
0=no limitation
1-100= (initial zero point) ±1%FS -
(initial zero point) ±100%FS
*If (REGULA) ≠none, the max is 2
AZSM
0-100
8
Zero tracking window
0=0d, no tracking
1-100=±(0.2+0.05*(1-100))d /s
*If (REGULA) ≠none, the max is 4
CONFIG
SubMenu1
SubMenu2
Option
Default
Parameter Description
Comment